Table 1:

PCD-CT versus EID-CT scan and reconstruction parameters

ParameterEIDPCD
Scanner modelForceAlpha
kV100/Sn150 kV120 kV (manual kV)
QRM/Effective mAs/IQ-levelHead/Neck CTA: 360/225 (AEC on), Head CTA: 280/175 (AEC off)230 (CAREkeV IQ Level)
Collimation (mm)192 × 0.6120 × 0.2
CTDIvol (mGy)37.1 ± 4.736.1 ± 4.0
Section/Increment (mm)0.6/0.4 (mixing ratio 0.5)0.6/0.4; 0.2/0.2
KernelQr54Qr56; Qr89
Matrix10241024
Iterative reconstruction and noise reductionADMIRE-3Quantum iterative reconstruction Strength 3; UHR-CNN
Contrast injectionOminipaque 350, 100 mL at 4 mL/s, followed by 35 mL saline at 4 mL/sOminipaque 350, 100 mL at 4 mL/s, followed by 35 mL saline at 4 mL/s
  • Note:—IQ indicates image quality (a PCD setting); AEC, automatic exposure control.