Table 2:

Acquisition parameters for MR imaging sequences in the MS protocol

SequencePlaneTR (ms)TE (ms)TI (ms)Image MatrixFOV (mm)Section (mm)Scan Time (min)
PSIRAxial430013400256 × 25624034.0
DIRAxial15,000253400/325512 × 51224036.5
Dual-FSEAxial680010/90256 × 25624036.0
FLAIRAxial10,000802600256 × 25624033.0
  • Note:—MS indicates multiple sclerosis; PSIR, phase-sensitive inversion recovery; DIR, double inversion recovery; FSE, fast spin-echo; FLAIR, fluid-attenuated inversion recovery; FOV, field of view.