Table 1: Parameters of SP-CTA, PCT, and MP-CTA protocol
Parameter SP-CTA PCT MP-CTA Detector number 64 64 64 Scanning mode Helical Cine Cine Coverage length 40.0 mm 40.0 mm 40.0 mm Reconstructed section thickness 0.625 mm 5.0 mm 0.625 mm Pitch 0.516:1 – – Speed 20.62 (mm/rot) – – Rotation time 0.4 second 1.0 second 1.0 second Coverage speed 51.55 mm/s – – No. of images per rotation – 8i 64i Cine time interval – 0.5 second 0.5 second kVp 120 80 80 mA 350 180 180 Contrast volume 60 mL 30 mL 30 mL CTDI 51.72 623.31 623.31 Scanning range No limitation 40 mm 40 mm Phase selection No Yes Yes
Note:—rot indicates rotation; –, not applicable; CTDI, CT dose index; SP-CTA, single-phase CT angiography; PCT, perfusion CT; MP-CTA, multiphase CT angiography.