Array ( [urn:ac.highwire.org:guest:identity] => Array ( [runtime-id] => urn:ac.highwire.org:guest:identity [type] => guest [service-id] => ajnr-ac.highwire.org [access-type] => Controlled [privilege] => Array ( [urn:ac.highwire.org:guest:privilege] => Array ( [runtime-id] => urn:ac.highwire.org:guest:privilege [type] => privilege-set [privilege-set] => GUEST ) ) [credentials] => Array ( [method] => guest ) ) ) 1nareshArray ( [urn:ac.highwire.org:guest:identity] => Array ( [runtime-id] => urn:ac.highwire.org:guest:identity [type] => guest [service-id] => ajnr-ac.highwire.org [access-type] => Controlled [privilege] => Array ( [urn:ac.highwire.org:guest:privilege] => Array ( [runtime-id] => urn:ac.highwire.org:guest:privilege [type] => privilege-set [privilege-set] => GUEST ) ) [credentials] => Array ( [method] => guest ) ) )Table 1:Parameters for carotid MR imaging protocols at 3T
Parameter T1WI With or Without Contrast Enhancement T2WI TOF Imaging CE-MRA Acquisition mode 2D 2D 3D 3D Acquisition sequence Fast spin-echo Fast spin-echo Spoiled gradient and flow compensation Enhanced gradient recalled-echo Blood suppression technique Quadruple inversion recovery Multisection double inversion recovery Saturation veins None TE (ms) 11 52 3.5 1.7 TR (ms) 800 4000 23 5.5 TI (ms) 520 250 n/a n/a Echo-train length 10 12 n/a n/a Excitation flip angle (degrees) 90 90 20 30 No. of signals acquired 1 2 1 FOVa 160 × 130 mm 160 × 120 mm 160 × 160 mm 150 mm Matrix size 256 × 256 256 × 256 288 × 256 256 × 256 No. of sections 18 18 48 40 Section thickness (mm) 2 2 1 0.8 Coverage (mm) 36 36 44 32 mm Imaging time (min:sec) 5:49 4:32 4:46 0:59
a The FOV of 140 × 140 cm was also used in some patients.