1naresh
Array ( [urn:ac.highwire.org:guest:identity] => Array ( [runtime-id] => urn:ac.highwire.org:guest:identity [type] => guest [service-id] => ajnr-ac.highwire.org [access-type] => Controlled [privilege] => Array ( [urn:ac.highwire.org:guest:privilege] => Array ( [runtime-id] => urn:ac.highwire.org:guest:privilege [type] => privilege-set [privilege-set] => GUEST ) ) [credentials] => Array ( [method] => guest ) ) ) 1nareshArray ( [urn:ac.highwire.org:guest:identity] => Array ( [runtime-id] => urn:ac.highwire.org:guest:identity [type] => guest [service-id] => ajnr-ac.highwire.org [access-type] => Controlled [privilege] => Array ( [urn:ac.highwire.org:guest:privilege] => Array ( [runtime-id] => urn:ac.highwire.org:guest:privilege [type] => privilege-set [privilege-set] => GUEST ) ) [credentials] => Array ( [method] => guest ) ) )Summary of 10 technical problems during GDC deployment
No. Sex/Age (y) Aneurysm Size H&H Event(s) GDC* Catheter Management Clinical Consequence 1 F/54 BA 8 0 SB, IL 6th and 7th coil/9, GDC-10, 2 × 6, 2 × 4, S, SR, SG Rebar14 Removal of the interlocked coils and the catheter as one unit Uneventful 2 F/53 BA 6 2 SB 2nd coil/6, GDC-10, 4 × 8, 2D, S, SR, SG Excelsior Reinsertion with subsequent coil Uneventful 3 F/48 ACoA 3 2 SB last coil/2, GDC-10, 3 × 6, 2D, S, SR, SG Rebar14 Reinsertion with subsequent coil Uneventful 4 F/61 ICA 10 0 SD, SB last coil/7, GDC-10, 3 × 6, 2D, S, SR, SG Excelsior Heparinization, antiplatelet therapy Uneventful 5 M/42 ACoA 5 4 SD, UD last coil/7, GDC-10, 2 × 1, S, SR, SG Rebar14 Heparinization, antiplatelet therapy Uneventful 6 M/46 ACoA 2 2 SD last coil/2, GDC-10, 2 × 1, US, SR, SG Excel14 Coil removal with snare, no further aneurysm treatment Uneventful 7 F/52 ICA 24 0 SD 1st coil/1, GDC-18, 18 × 30, 2D, SG Excelsior Urgent bypass surgery and subsequent endovascular occlusion of the artery Uneventful 8 F/58 BA 16 4 SD 1st coil/16, GDC-10, 8 × 20, 3D, SG Excelsior No special management Uneventful 9 F/66 PICA 3 2 UD last coil/3, GDC-10, 2 × 2, S, SR, SG Rebar14 Heparinization, antiplatelet therapy Uneventful 10 F/43 ACoA 3 4 UD last coil/2, GDC-10, 3 × 6, S, SR, SG Rebar14 Antiplatelet therapy Uneventful
Note.—H&H indicates Hunt and Hess grade of subarachnoid hemorrhage; size, the largest diameter of the aneurysm (mm); BA, basilar top aneurysm; SB, slipping-back of the coil; IL, coil interlocking; S, soft-GDC; SR, stretch-resistant-GDC; SG, synerG-GDC; 2D, 2-diameter GDC; 3D, 3D GDC; ACoA, anterior communicating artery aneurysm; ICA, carotid artery aneurysm; SD, spontaneous detachment; UD, undesirable detachment at the parent artery; PICA, posterior inferior cerebellar artery aneurysm.
* GDC coil(s) directly related to the event(s)/total number of GDCs inserted. Helical diameter (mm) × length (cm).