Table 1: Maximum entrance air kerma rates for the 3 fluoroscopic modes I, II, and III on the reference and dose-reduction platformsa
Platform, Mode Frames/Second Filtration Maximum EAKR (mGy/min) Reference I 6 0.4 mm Cu +1 mm Al 22 II 12.5 0.4 mm Cu +1 mm Al 44 III 12.5 0.1 mm Cu +1 mm Al 79 DRT I 15 0.4 mm Cu +1 mm Al 11 II 15 0.4 mm Cu +1 mm Al 26 III 15 0.1 mm Cu +1 mm Al 62
Note:—DRT indicates dose-reduction technology.
↵a The operator chooses the fluoroscopy mode on the dose-reduction platform independent of the DSA program and the targeted dose-reduction setting used for angiography.