Table 1:

Technical parameters used for radiofrequency ablation, expected minor-axis diameter of the ablation zones based on nonanimal models, and measured ablation zone diameters on MRI and gross pathology

WattsThermocouple Temperatures (°C)Ablation Time (sec)Diameter of Ablation Zone Orthogonal to Ablation Probe (mm)
ProximalDistalExpectedMRIGross Pathology
1072519220.013.413.8
5584210716.69.210.0
551404213.07.38.0